PCE-NDL 10 is a high-quality noise dose meter or sound level meter with data logging capabilities. This noise dosimeter records sound pressure level (SPL) measurements as well as dosimetry (dose) measurements in decibels (dB) and saves the measurement data to the internal memory, which can then be transferred to an SD card.
Note: To set up the SD card for the first time, the PCE-NDL 10 must be connected to AC power adapter (AC power supply NET-LUT or NET-LUT-US sold separately - see accessories tab).
Sound Level Meter Specifications | |
---|---|
Type | SPL and dose |
Measuring range | 30 ... 130 dB (SPL) / 70 ... 130 dB (dose, in 1 dB increments) |
Dose rating levels | 80, 84, 85, 90 dB |
Dose exchange rate | 3, 4, 5 or 6 dB |
Accuracy (at 23 ± 5 °C, < 94 dB) |
± 3.5 dB (31.5 Hz) ± 2.5 dB (63 Hz) ± 2.0 dB (125 Hz) ± 1.9 dB (250 Hz) ± 1.9 dB (500 Hz) ± 1.4 dB (1000 Hz) ± 2.6 dB (2000 Hz) ± 3.6 dB (4000 Hz) ± 5.6 dB (8000 Hz) |
Resolution | 0.1 dB |
Frequency range | 31.5 ... 8000 Hz |
Frequency weighting | A and C |
Time evaluation | 125 ms (fast) / 1 s (slow) |
Data logging | Manual (interval must be set to "0") or adjustable automatic |
Adjustable auto logging interval | 1 ... 3600 s |
Sampling rate | Approx. 1 second |
Storage | Internal memory for up to 16,000 readings, SD card slot for 1 ... 4GB SD card (card included) |
Interface | RS-232 and USB |
Measurement error | ≤ 0.1% of the total stored data |
Measurement outside range | Display of "----" |
Functions | Data hold, min / max, date / time, auto shutdown (on / off), audible alarm (on / off) |
Calibration | Internal calibration possible at 94 dB with PCE-SC 42 acoustic calibrator (sold separately) |
Power supply | 6 AAA Batteries or optional power supply |
Operating conditions | 0 ... +50 °C (+32 ... +122 °F), < 85% RH |
Display | Backlit LCD, 50 x 30 mm (1.97 x 1.18") |
Microphone | Electric condensation / Ø 12.7 mm (0.5") |
Approx. device weight | 250 g (0.56 lb) |
Approx. device dimensions | 132 x 80 x 32 mm (5.20 x 3.15 x 1.26") |